Sartorius Basic lite Instrukcje Operacyjne Strona 115

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 132
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 114
115
k ) Activate menu item 1-19-2.
Enter the resolution (lowest scale interval d) (For example 0.01 lb). This value must
corresponed with the value entered under menu item 11-4-1 (and/or 11-5-1 or 11-6-1).
) Confirms setting.
k ) Activate menu item 1-19-3.
Enter the sensitivity of the load cells in mV/V (obtain data from the data sheet, for example).
On a weighing platform that consists of several load cells, the sensitivity of load cell 1)
should be entered here.
Example: The sensitivity of the load cell is 1.9440 mV/V.
) Confirms setting.
k ) On a weighing platform that consists of several load cells, enter the sensitivity of
the other load cells (max. 4 load cells) under the menu items 1-19-4 (for cell 2) to
k ) 1-19-6 (for cell 4). If there are less than 4 load cells for the corresponding menu items,
enter the value “0.000".
k )
k )
Call up menu item 1-19-7 (save values entered on menu page 1-19).
) ( ( Save ID codes of the load cell(s), return to menu page “1."
Set/delete preload: see corresponding section on the subsequent pages.
See also chapter entitled “Operating the Combics", under “Calibration, Adjustment"
or the Service Manual.
Restart: Switch the terminal off and on again.
After completing the A/D converter configuration, move the menu access switch from the
“open” to the “locked” position, as follows:
§ Remove the cover plate from the back of the indicator housing
§ Move the switch to the left (“locked”).
See also the “Calibration/Adjustment” section in the chapter entitled
“Operating the Combics”, or refer to the service manual.
e e Turn the scale off and on again to restart it.
After the self-test, the scale is ready to operate.
Przeglądanie stron 114
1 2 ... 110 111 112 113 114 115 116 117 118 119 120 ... 131 132

Komentarze do niniejszej Instrukcji

Brak uwag